CareFusion Provides Update On Voluntary Global Recall Of AVEA Ventilator

CareFusion Provides Update on Voluntary Global Recall of AVEA® Ventilator

SAN DIEGO, May 27, 2015 /PRNewswire/ -- CareFusion, a BD company (NYSE:BDX), provided an update on a global voluntary recall that was initiated on April 21, 2015 to address an issue with certain units of AVEA® ventilators.

The recall has been designated as a Class I recall by the U.S. Food and Drug Administration (FDA), which means the Agency believes that there is a reasonable probability that use of the recalled product will cause serious adverse health consequences or death. To date, no report of patient injury has been received related to this issue.

The AVEA ventilator is only used in hospitals and other health care facilities and is intended for continuous breathing support for neonatal through adult patients. The recall is in response to a potential malfunction of an AVEA ventilator specific 5 psi pressure transducer. The affected AVEA ventilators may develop a failure mode over a period of time, where, by design, the ventilator activates false Extended High Ppeak or Circuit Occlusion audio and visual alarms, opens the safety valve and stops ventilating. If this occurs, alternate ventilation support will be required to reduce the potential of hypoxemia or hypercapnia.

The global recall involves AVEA ventilators manufactured, serviced and distributed from July 1, 2011 to March 15, 2015. CareFusion learned of this issue through customer reports identifying Extended High Ppeak or Circuit Occlusion alarms. A list of affected model and serial numbers is available at the CareFusion website. Refer to Figure 1 for the location of model and serial numbers.

Figure 1: Rear view of AVEA ventilator.

CareFusion puts patient safety first and will continue to work diligently with customers and global regulatory authorities to resolve this issue in a timely manner. The company has notified customers of the recall with an urgent recall letter. The CareFusion Recall Support Center is contacting global customers via telephone to coordinate on-site AVEA ventilator correction in a timely, effective manner.

The FDA and other regulatory authorities have been notified.

Instructions to Customers

CareFusion does not require the return of affected AVEA ventilators.

In the interim, if an AVEA ventilator exhibits a sustained Extended High Ppeak or Circuit Occlusion alarm followed by the opening of the safety valve that cannot be cleared by powering the ventilator off and back on again, immediately remove the ventilator from service, provide alternate ventilation and contact CareFusion Technical Support at the contact information listed below to report the issue.

Customer inquiries related to this action should be addressed to the CareFusion Recall Support Center at 1.888.562.6018 or email SupportCenter@carefusion.com between the hours of 6:30 a.m. to 5:00 p.m. PDT.

For technical support or to report a problem, please contact CareFusion Technical Support at 1.800.231.2466 or email Support.Vent.US@carefusion.com.

Adverse reactions or quality problems experienced with the use of this product should be reported to the FDA’s MedWatch Adverse Event Reporting program either online, by regular mail or by fax:

Countries with AVEA ventilators affected by the global recall:

ISO Code

Country

AL

Albania

AS

American Samoa

AR

Argentina

AU

Australia

AT

Austria

BH

Bahrain

BD

Bangladesh

BB

Barbados

BE

Belgium

BZ

Belize

BM

Bermuda

BO

Bolivia

BA

Bosnia and Heherzegowina

BR

Brazil

CA

Canada

CL

Chile

CN

China (People’s Rep)

CO

Colombia

CR

Costa Rica

HR

Croatia

CY

Cyprus

CZ

Czech (Rep)

DO

Dominican Republic

EC

Ecuador

EG

Egypt

SV

El Salvador

EE

Estonia

FR

France

GE

Georgia

DE

Germany

GB

Great Britain

GR

Greece

GT

Guatemala

HN

Honduras (Rep)

HK

Hong Kong, China

HU

Hungary (Rep)

IN

India

ID

Indonesia

IE

Ireland

IL

Israel

IT

Italy

JP

Japan

JO

Jordan

KR

Korea (Rep)

KW

Kuwait

LV

Latvia

LB

Lebanon

LY

Libyan Jamahiriya

LT

Lithuania

MY

Malaysia

MX

Mexico

MD

Moldova (republic of)

KR

Korea (Rep)

KW

Kuwait

MA

Morocco

NP

Nepal

NL

Netherlands

Nl

Nicaragua

NO

Norway

OM

Oman

PK

Pakistan

PS

Palestine

PA

Panama (Rep)

PY

Paraguay

PE

Peru

PH

Philippines

PL

Poland

PT

Portugal

PR

Puerto Rico

QA

Qatar

RO

Romania

RU

Russian Federation

RW

Rwanda

SA

Saudi Arabia

SG

Singapore

SK

Slovakia

SI

Slovenia

ZA

South Africa

ES

Spain

TW

Taiwan

TH

Thailand

TN

Tunisia

TR

Turkey

AE

United Arab Emirates

UK

United Kingdom

UY

Uruguay

US

United States

VE

Venezuela

VN

Vietnam

YE

Yemen

About CareFusion
CareFusion, a BD company, serves the health care industry with products and services that help hospitals measurably improve the safety and quality of care. The company develops industry-leading technologies including Alaris® infusion pumps and IV sets, MaxPlus® and MaxZero IV connectors and sets, Pyxis® automated dispensing and patient identification systems, AVEA®, LTV® series and AirLife® ventilation and respiratory products, ChloraPrep® products, MedMined® services for data mining surveillance, V. Mueller® surgical instruments, and an extensive line of products that support interventional medicine. For more information please visit www.carefusion.com.

About BD
BD is a leading medical technology company that partners with customers and stakeholders to address many of the world’s most pressing and evolving health needs. Our innovative solutions are focused on improving medication management and patient safety; supporting infection prevention practices; equipping surgical and interventional procedures; improving drug delivery; aiding anesthesiology and respiratory care; advancing cellular research and applications; enhancing the diagnosis of infectious diseases and cancers; and supporting the management of diabetes. We are more than 45,000 associates in 50 countries who strive to fulfill our purpose of “Helping all people live healthy lives” by advancing the quality, accessibility, safety and affordability of healthcare around the world. In 2015, BD welcomed CareFusion and its products into the BD family of solutions. For more information on BD, please visit www.bd.com.

Photo - http://photos.prnewswire.com/prnh/20150526/218576

To view the original version on PR Newswire, visit:http://www.prnewswire.com/news-releases/carefusion-provides-update-on-voluntary-global-recall-of-avea-ventilator-300089097.html

SOURCE CareFusion

Help employers find you! Check out all the jobs and post your resume.

MORE ON THIS TOPIC