Covidien (NYSE:COV) issued a field safety alert on some of its Medi-Trace Cadence and Kendall defibrillation electrodes, saying a connector compatibility issue with defibrillators made by Philips (NYSE:PHG) may have contributed to the death of a patient.
The electrodes in question will not connect with Philips FR3 or FRx AED defibrillators and can result in a delay of therapy, according to a press release. Covidien said it received 2 reports in which customers attempted to use a Covidien electrode with a Philips FR3 AED unit, with the mismatch contributing to "a delay in resuscitation" that "may have contributed to the subsequent death of 1 patient."
Help employers find you! Check out all the jobs and post your resume.
The electrodes in question will not connect with Philips FR3 or FRx AED defibrillators and can result in a delay of therapy, according to a press release. Covidien said it received 2 reports in which customers attempted to use a Covidien electrode with a Philips FR3 AED unit, with the mismatch contributing to "a delay in resuscitation" that "may have contributed to the subsequent death of 1 patient."