Performance of the PointCare NOW System for CD4 Counting in HIV Patients Based on Five Independent Evaluations
Published: Thursday, August 09, 2012
Author: Michèle Bergeron et al.

by Michèle Bergeron, Géraldine Daneau, Tao Ding, Nadia E. Sitoe, Larry E. Westerman, Jocelijn Stokx, Ilesh V. Jani, Lindi M. Coetzee, Lesley Scott, Anja De Weggheleire, Luc Boel, Wendy S. Stevens, Deborah K. Glencross, Trevor F. Peter

Introduction

Point-of-care (POC) CD4 testing can improve access to treatment by enabling decentralization and reducing patient loss-to-follow-up. As new POC CD4 technologies become available, their performance should be assessed before widespread deployment. This study reports the findings of five independent evaluations of the PointCare NOW CD4 system.

Materials/Methods

Evaluations were conducted in Southern Africa (Mozambique, South Africa) and North America (Canada, USA). 492 blood samples (55 from HIV-negative blood donors and 437 from HIV-infected patients, including 20 children aged between 12 and 59 months) were tested with both the PointCare NOW and reference flow cytometry instruments. Assessment of bias, precision and levels of clinical misclassification for absolute and percent CD4 count was conducted.

Results

PointCare NOW significantly overestimated CD4 absolute counts with a mean relative bias of +35.0%. Bias was greater in samples with CD4 counts below =350cells/µl (+51.3%) than in the CD4 >350cells/µl stratum (15.1%). Bias in CD4% had a similar trend with an overall relative mean bias of +25.6% and a larger bias for low CD4 stratum (+40.2%) than the higher CD4 stratum (+5.8%). Relative bias for CD4% in children was -6.8%. In terms of repeatability, PointCare NOW had a coefficient of variation of 11%. Using a threshold of 350cells/µl, only 47% of patients who qualified for antiretroviral therapy with reference CD4 testing, would have been eligible for treatment with PointCare NOW test results. This was 39% using a 200cells/µl threshold. Agreement with infant samples was higher, with 90% qualifying at a 25% eligibility threshold.

Conclusion

The performance of the PointCare NOW instrument for absolute and percent CD4 enumeration was inadequate for HIV clinical management in adults. In children, the small sample size was not large enough to draw a conclusion. This study also highlights the importance of independent evaluation of new diagnostic technology platforms before deployment.