BioSpace.com

Biotech and Pharmaceutical
News & Jobs
Search the Site
 
   
Biotechnology and Pharmaceutical Channel Medical Device and Diagnostics Channel Clinical Research Channel BioSpace Collaborative    Job Seekers:  Register | Login          Employers:  Register | Login  

NEWSLETTERS
Free Newsletters
Archive
My Subscriptions

NEWS
News by Subject
News by Disease
News by Date
PLoS
Search News
Post Your News
JoVE

CAREER NETWORK
Job Seeker Login
Most Recent Jobs
Browse Biotech Jobs
Search Jobs
Post Resume
Career Fairs
Career Resources
For Employers

HOTBEDS
Regional News
US & Canada
  Biotech Bay
  Biotech Beach
  Genetown
  Pharm Country
  BioCapital
  BioMidwest
  Bio NC
  BioForest
  Southern Pharm
  BioCanada East
  US Device
Europe
Asia

DIVERSITY

INVESTOR
Market Summary
News
IPOs

PROFILES
Company Profiles

START UPS
Companies
Events

INTELLIGENCE
Research Store

INDUSTRY EVENTS
Biotech Events
Post an Event
RESOURCES
Real Estate
Business Opportunities

 News | News By Subject | News by Disease News By Date | Search News
eNewsletter Signup
Miles
Km80.5

   

Zimmer Spine, Inc. Issues Worldwide Recall of all 315 Units of the PEEK Ardis® Inserter


12/20/2012 7:18:24 AM

MINNEAPOLIS, Dec. 20, 2012 /PRNewswire/ -- Zimmer Spine, Inc. is initiating a voluntary worldwide recall of all 315 units of the PEEK Ardis® Inserter. The Ardis Inserter is a surgical instrument used during spinal surgery to implant the PEEK Ardis Interbody Spacer. The inserters are being recalled because Zimmer Spine has received reports of PEEK Ardis Interbody Spacer implant breakage when the implant is subjected to excessive lateral and/or off-axis forces from the inserter during surgery. Intra-operative complaint reports received to date indicate an occurrence rate of 0.52%. Observed health risks associated with implant fragments may include dural tears and blood loss. No post-operative complaints have been reported that are attributed to a fractured implant. Patients with concern about their implant should contact their healthcare provider.

(Photo: http://photos.prnewswire.com/prnh/20121220/DE33015 )

Surgeons and hospitals with PEEK Ardis Inserter instruments should immediately stop using the inserters and return them to Zimmer Spine. Surgery using the PEEK Ardis Interbody Spacer cannot be performed as intended without the aid of the PEEK Ardis Inserter instrument.

This recall includes all lots of the following part numbers that were distributed from June 2008 through December 2012:

Inserter in Ardis Instrument Set

3256-01

Specialty Ardis Inserter

07.00780.701

650-0176-01

650-0239-01

650-0297-01

07.00780.702

650-0179-01

650-0263-01

650-0351-01

07.00900.701

650-0208-01

650-0293-01

650-0351-02

07.00900.702



650-0365-01

Products can be identified by the part number and lot number physically marked on the surgical instrument. Products were distributed nationwide in the United States and around the world. Zimmer Spine has notified the FDA of this action and is working with other regulatory agencies worldwide.

Zimmer Spine is notifying its distributors and customers worldwide via a letter and is arranging for the return of all PEEK Ardis Inserters.

Further information, including a detailed listing of part and lot numbers, is available at www.peekardis.zimmer.com.

Healthcare professionals or customers may report adverse events or quality problems experienced with the use of these products to Zimmer Spine by calling 866-774-6368 between the hours of 8AM and 5PM CST or by email at qualitycompliance@zimmer.com.

Adverse reactions or quality problems experienced with the use of these products may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

SOURCE Zimmer Spine, Inc.