BioSpace.com

Biotech and Pharmaceutical
News & Jobs
Search the Site
 
   
Biotechnology and Pharmaceutical Channel Medical Device and Diagnostics Channel Clinical Research Channel BioSpace Collaborative    Job Seekers:  Register | Login          Employers:  Register | Login  

NEWSLETTERS
Free Newsletters
Archive
My Subscriptions

NEWS
News by Subject
News by Disease
News by Date
PLoS
Search News
Post Your News
JoVE

CAREER NETWORK
Job Seeker Login
Most Recent Jobs
Browse Biotech Jobs
Search Jobs
Post Resume
Career Fairs
Career Resources
For Employers

HOTBEDS
Regional News
US & Canada
  Biotech Bay
  Biotech Beach
  Genetown
  Pharm Country
  BioCapital
  BioMidwest
  Bio NC
  BioForest
  Southern Pharm
  BioCanada East
  US Device
Europe
Asia

DIVERSITY

INVESTOR
Market Summary
News
IPOs

PROFILES
Company Profiles

START UPS
Companies
Events

INTELLIGENCE
Research Store

INDUSTRY EVENTS
Biotech Events
Post an Event
RESOURCES
Real Estate
Business Opportunities

 News | News By Subject | News by Disease News By Date | Search News
eNewsletter Signup
Miles
Km80.5

   

Respironics, Inc. (RESP) Ventilators Recalled for Software Malfunctions


6/21/2013 8:47:24 AM

Staying up-to-date has never been simpler. Sign up for the free GenePool newsletter today!

Respironics California, Inc., V60 Ventilators

Recall Class: Class I

Date Recall Initiated: June 4, 2013

Product: Respironics California, Inc., V60 Ventilators

To find Respironics Material P/N (Philips 12 Digit P/N) and Serial Numbers that are affected by this recall please see: http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?ID=118739

The recalled products were manufactured from March, 2009 through April, 2013 and distributed from November 16, 2009 through March 28, 2013.

Use: The Respironics V60 ventilator provides continuous or intermittent breathing support to pediatric patients weighing 44 pounds or greater to adult patients. This ventilator is used in hospitals or other health care facilities under the direction of qualified medical professionals such as physicians, nurses, and respiratory therapists.

Recalling Firm:

Respironics California, Inc.

2271 Cosmos Court

Carlsbad, California 92011-1517

Reason for Recall: Respironics California, Inc. has initiated a recall on the V60 ventilator because of an issue with the software on the V60 Power Management Board Assembly. If a component fails on the Power Management Board Assembly, it may cause ventilator support to be lost with potentially no audible alarm from the ventilator. This recalled product may cause serious adverse health consequences, including death.

Public Contact: Customers may contact their local Philips Respironics representative at 1- 800- 722-9377 for any questions about this recall.

FDA District: Los Angeles District Office

FDA Comments:

On June 3, 2013, Philips Healthcare sent a "MEDICAL DEVICE CORRECTION" letter to all customers who received the V60 Ventilators. The letter described the product, the problem, and the actions to be taken.

Customers were informed that the V60 ventilator may continue to be used according to its directions for use, pending the completion of the software update. In addition, customers were instructed to refer to their manual for additional information on warnings.

A Philips Field Service Engineer, Approved Service Provider, or Distributor will be contacting customers to schedule a no-cost update and replacement of the Power Management Board Assembly Software on all V60 ventilators shipped from the manufacturer prior to April 1, 2013.

Customers were instructed to contact their local Philips Respironics representative at 1-800-722-9377 for any further information or support concerning this issue.

Class I recalls are the most serious type of recall and involve situations in which there is a reasonable probability that use of these products will cause serious adverse health consequences or death. Health care professionals and consumers may report adverse reactions or quality problems they experienced using these products to MedWatch: The FDA Safety Information and Adverse Event Reporting Program either online, by regular mail or by FAX.

Help employers find you! Check out all the jobs and post your resume.