BioSpace.com

Biotech and Pharmaceutical
News & Jobs
Search the Site
 
   
Biotechnology and Pharmaceutical Channel Medical Device and Diagnostics Channel Clinical Research Channel BioSpace Collaborative    Job Seekers:  Register | Login          Employers:  Register | Login  

NEWSLETTERS
Free Newsletters
Archive
My Subscriptions

NEWS
News by Subject
News by Disease
News by Date
PLoS
Search News
Post Your News
JoVE

CAREER NETWORK
Job Seeker Login
Most Recent Jobs
Browse Biotech Jobs
Search Jobs
Post Resume
Career Fairs
Career Resources
For Employers

HOTBEDS
Regional News
US & Canada
  Biotech Bay
  Biotech Beach
  Genetown
  Pharm Country
  BioCapital
  BioMidwest
  Bio NC
  BioForest
  Southern Pharm
  BioCanada East
  US Device
Europe
Asia

DIVERSITY

INVESTOR
Market Summary
News
IPOs

PROFILES
Company Profiles

START UPS
Companies
Events

INTELLIGENCE
Research Store

INDUSTRY EVENTS
Biotech Events
Post an Event
RESOURCES
Real Estate
Business Opportunities

 News | News By Subject | News by Disease News By Date | Search News
eNewsletter Signup
Miles
Km80.5

   

Hamilton Medical, Inc. Recalls Ventilators Over Software Glitch


2/7/2013 10:08:00 AM

February 07, 2013 -- Recall Class: Class I

Date Recall Initiated: January 11, 2013

Product: HAMILTON-T1 ventilators with software versions 1.1.2 and lower; part numbers 161005 and 161006

This device was manufactured and distributed from February, 2012 through December, 2012.

Use: The HAMILTON-T1 ventilator provides continuous or intermittent breathing support to adults and pediatric patients.

Recalling Firm:

Hamilton Medical Inc.

4990 Energy Way

PO Box 30008

Reno, Nevada 89502-4123

Manufacturer:

Hamilton Medical AG

Via Crusch 8

Bonaduz, Switzerland

Reason for Recall: During ventilation of small pediatric patients with high airway resistance and low lung function, there may be unexpected high internal oxygen consumption by HAMILTON-T1 ventilators with software versions 1.1.2 and lower. This may cause miscalculation of the required oxygen needed for long time applications with limited oxygen supply. The HAMILTON-T1 oxygen capacity must be calculated using a larger margin than originally expected.

This product may cause serious adverse health consequences, including death.

Public Contact: Customers with questions may call the company at 1-800-426-6331, extension 215.

FDA District: San Francisco

FDA Comments:

On January 11, 2013, Hamilton Medical mailed a letter notifying their customers of the problem and informed them that the firm is working on a software version with reduced oxygen consumption. The letter included a "Medical Device Safety Alert and Corrective Action" document. The firm asked their customers to review this document for details regarding the root cause and corrective action with the current 1.1.2 or lower software.

Customers will be notified upon completion of the new software. Arrangements will be made to provide a loaner unit while the customer’s T1 ventilator is returned to the firm’s Reno Service Department for the software upgrade.

The firm also directed their customers to update their T1 Operator Manuals with an included Oxygen Consumption formula page which provides revised information for calculating oxygen capacity.

Class I recalls are the most serious type of recall and involve situations in which there is a reasonable probability that use of these products will cause serious adverse health consequences or death.

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these products to MedWatch: The FDA Safety Information and Adverse Event Reporting Program either online, by regular mail or by FAX.



Read at BioSpace.com

   

ADD TO DEL.ICIO.US    ADD TO DIGG    ADD TO FURL    ADD TO STUMBLEUPON    ADD TO TECHNORATI FAVORITES